Technical data

ZEFIRO LC/MS DUAL PURITY ZEFIRO LC/MS DUAL PURITY S
STANDARD NITROGEN FLOW RATE LC/MS N2: 35 or 65 l/min at STP
Collision Cell N2: 200 ml/min at STP
 LC/MS N2: up to 35 l/min at STP
Collision Cell N2: 1100 ml/min at STP
OUTLET PRESSURE 7 bar at STP (102 psi)
2,5 bar at STP (36 psi)
7,2 bar at STP (105 psi)
6,5 bar at STP (95 psi)
STANDARD PURITY 99.0% at STP
99,999% at STP
98.0% at STP
99,999% at STP
INPUT VOLTAGE 230 V – 50 or 60 Hz 230 V – 50 or 60 Hz
WEIGHT 130 Kg 125 Kg
POWER CONSUMPTION 1800 or 1900 W 1900 W
PRESSURE ACCURACY 0.1 bar (± 0.5 %) 0.1 bar (± 0.5 %)
MICROPROCESSOR CONTROLLED DISPLAY Graphic display, 128 x 64 px Graphic display, 128 x 64 px
INDEX OF PROTECTION IP20 IP20
TEMPERATURE 5°C to +30°C 5°C to +30°C
RELATIVE HUMIDITY 0-80%, non condensing 0-80%, non condensing
SOUND PRESSURE LEVEL 50 or 55 dB(A) 55 dB(A)
OUTPUT PORT G 1/4 + quick connection for 1/4’’ LC/MS
G 1/4 + 1/8’’ swagelock for Collision Cell
G 1/4 + quick connection for 1/4’’ LC/MS
G 1/4 + 1/8’’ swagelock for Collision Cell
CASE DIMENSIONS width 48 cm, height 70 cm,
length 84 cm
width 48 cm, height
70 cm, length 84 cm

STP: Standard Temperature and Pressure (20°C, 1 bar)